Chongqing Bio Newvision Medical Equipment Ltd.

Build Ophthalmic Equipment Flag Exploit Dignosis New Era

Manufacturer from China
Verified Supplier
4 Years
Home / Products / Automated Perimetry Machine /

Fuse Type Automated Perimetry Test Machine 240V Multiple Report Analysis

Contact Now
Chongqing Bio Newvision Medical Equipment Ltd.
Visit Website
City:chongqing
Province/State:chongqing
Country/Region:china
Contact Person:MrRobert Wang
Contact Now

Fuse Type Automated Perimetry Test Machine 240V Multiple Report Analysis

Ask Latest Price
Video Channel
Brand Name :SYSEYE
Model Number :BIO-1000
Certification :CE
Place of Origin :China
MOQ :1
Price :$2,800.00-$3,200.00
Payment Terms :T/T
Supply Ability :`100
Delivery Time :5 working days
Packaging Details :412mm ( length) * 497mm ( width) * 513mm ( height )
Product name :Provide multiple report analysis Automatic Computer Perimeter
Input power of the equipment :300 VA
Fuse element :fuse type T2AL250V
Power :100- -240V,50-60Hz
Feature :Provide multiple report analysis
Classified by operation mode :intermittent loading and continuous operation
Inspection range :90°
Inspection distance :30cm
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

peripheral vision test machine beautiful software interface multiple report analysis

BIO-1000 automatic computer vision instrument is developed by our R&D team to absorb the advantages of international advanced models. It integrates sound, light, mechanical and electrical technologies, adopts international general random detection mode for visual marking, and has a wide examination range (0-90°) to help doctors to judge the patient's eye condition more comprehensively.Automated Perimetry Machine Adopt the high-end computer configuration, compared with the domestic similar products, stability, reliability is higher.Support Windows XP, Win7 platform, beautiful software interface, simple operation, easy to learn and use.According to international standards, it provides a very comprehensive software testing program and multiple report analysis to assist in the diagnosis of diseases related to visual impairment and help to make a more accurate diagnosis

Automated Perimetry Machine Specifications:

Basic specification

Provide multiple report analysis ​ Automated Perimetry Machine

Inspection range

90°

Inspection distance

30cm

Background light brightness

White 31.5asb

Visual target brightness

1asb--- -10000asb

Visual target size

GoldmannIII

Visual target interval time

200ms

Visual target interval time

Standard or adjustment according to patient reflection

Threshold test model

Center5-°16, center10° -68, 24 -2, 30-2, 30° -60° nasal step

Upper Threshold test model

30°- 40, 30°-76, P-60, 60°-81, 60°-120,190°-135 nasal side

Special detection strategy

Esterman monocular, Esterman binocular . user- customized test, 150P driver monocular fast detection, 150P driver monocular standard detection , upper 36 detection, blind spot detection,150horizontal straightness detection

Pupil size test

Automatic

Fixation monitoring

Blind spot monitoring , eye position eye position deviation alarm, eye deviation curve

Environment light detection

Automatic

Analysis software

Reliability analysis , single vision report, triple vision report, GPA half vision analysis, GPA glaucoma developing analysis , VFI vision index analysis

Operation system

Win7 above

Feature

Automated Perimetry Machine
1. Comprehensive real-time monitoring, heiji-krakau physiological blind spot monitoring. Gaze tracking/head position tracking, automatic measurement of pupil diameter, reduce the impact of pupil effect on visual field detection.
2. Personalized design, accurate clinical analysis, accurate and rapid examination strategy.
3. Under international Goldman standard, providing a variety of calssic test procedures and report analysis.



Background brightness and test stimulus brightness of the product are measured at the position of the pupil of the patient's eye.
1.1 background brightness
Background light: white with backlight intensity of 10 CD /m2.
1.2 test stimuli are within the tolerance range specified in table 1.

1Parameter of stimulus

NO.ContentTolerance
1Background light,+25%,-20%
2Contrast ,Background white,stimulus whiteTolerance +25%,-20%
3Stimulus position0°~10°:≤0.5°
10°~30°:≤1°
>30°:≤2°
4Stimulus sizePoint angle:+20%,-15%
5Interval time200ms, ±20%
6Scope of backgroundMost of the peripheral stimulus points do not exceed 2° beyond the boundary
7Stimulus shapeThe ratio of the short axis length to the long axis length of the stimulus points is shown in table 2


























Inquiry Cart 0